Description The system serves to de-aerate encapsulated heating elements (products). Operating programs can be compiled and managed by means of a software (web visualization).For each product a special program can be defined. A program consists of maximum 20 steps. One step comprises pressure, time, and status of the vibration table. Operation of the system is done via touch screen. Basic Equipment: Vibration system consisting of chamber with built-in vibration table, control cabinet with integrated electronics, as well as signaling column, vacuum pump, valves, pressure sensor, and door mechanism.
Details Chamber volume: approx. 350 l Max. load: 25 kg Frequency of vibration table [min-1]: 2.096 - 3.336 min-1 Minimum pressure, absolute [mbar]: 400 mbar Maximum pressure, absolute [mbar]: Ambient pressure (usually 960 – 1030 mbar), no overpressure above ambient pressure Pressure range of vibration table: 2 - 6 bar Compressed air: 5 bar (clean compressed air, filter 5µm) Air consumption [l/min]: 128 - 564 l/min Order No.: 2390 000 PDF Downloads: > Vacuum Vibration System Technical specifications Electrical supply: 230 V / 50 Hz (different voltage on request) Power input: 4.8 A Fuse: Automatic fuse set to 6 A Ambient temperature: 10°C - 40°C Dimensions (w x d x h): 960 x 800 x 1.300 mm Weight: approx. 200 kg